Accès libre

Stability Studies of Torque Transducers

À propos de cet article

Citez

The present paper discusses the stability studies carried out on a small number of torque transducers over the years with capacities ranging from 10 Nm to 1000 Nm. The torque transducers have been calibrated using the torque standard machine based on the written standard calibration procedure. The study reveals that the uncertainty of measurement of torque transducers has been varying appreciably and it is more for lower range. Besides, the deviation from their average values has also been studied and found to be varying through the years.

eISSN:
1335-8871
Langue:
Anglais
Périodicité:
6 fois par an
Sujets de la revue:
Engineering, Electrical Engineering, Control Engineering, Metrology and Testing