Results of the validation of the analytical method for the determination of tylosin in feeds

ParameterFortification level (mg kg-1)Results
Correlation coefficient0.9985
Linearity (working range), mg kg-10.05-2.0
LOD, mg kg-10.035
LOQ, mg kg-10.05
CCα, mg kg-10.085
CCβ, mg kg-10.091
Recovery, %(n = 18)0.0588.2
0.2578.9
2.0108.3
Repeatability, CV % (n = 6)0.052.4
0.259.9
2.03.0
Within-laboratoryreproducibility, CV %(n = 18)0.056.5
0.256.9
2.06.5
Uncertainty (k = 2, p < 95), %0.0521.4
0.2519.7
2.023.8