Open Access

Signal Detection of Multi-Channel Capillary Electrophoresis Chip Based on CCD


Cite

A kind of multi-channel capillary electrophoresis (CE) chip signal detection system based on CCD was developed. The output signal of the CCD sensor was processed by a series of pre-processing circuits and ADC, and then it was collected by the Field Programmable Gate Array (FPGA) chip which communicated with a host computer. The core in FPGA was designed to control the signal flow of the CCD and transfer the data to PC based on a Nios II embedded soft-processor. The application of PC was used to store the data and demonstrate the curve. The measurement of the fluorescent signals for different concentration Rhodamine B dyes is presented and the comparison with other detection systems is also discussed.

eISSN:
1335-8871
Language:
English
Publication timeframe:
6 times per year
Journal Subjects:
Engineering, Electrical Engineering, Control Engineering, Metrology and Testing