Cite

Figure 1

(A) QC phantom fixed at the breast position on the patient support. The upper portion is uniformly filled with water, while the radiographic details are visible in the lower portion. (B) Arrangement of the details at the bottom of the QC phantom. The rods are made of polyethylene (PE), nylon, polyoxymethylene (POM), polytetrafluoroethylene (PTFE), and BR12, a plastic material mimicking the attenuation properties of breast tissue. The structure of the low contrast PMMA insert, exhibiting the five holes of varying diameter, is clearly perceivable.
(A) QC phantom fixed at the breast position on the patient support. The upper portion is uniformly filled with water, while the radiographic details are visible in the lower portion. (B) Arrangement of the details at the bottom of the QC phantom. The rods are made of polyethylene (PE), nylon, polyoxymethylene (POM), polytetrafluoroethylene (PTFE), and BR12, a plastic material mimicking the attenuation properties of breast tissue. The structure of the low contrast PMMA insert, exhibiting the five holes of varying diameter, is clearly perceivable.

Figure 2

Example of reconstructed water image (left panel) and detail image (right panel), acquired at 38 keV and high dose.
Example of reconstructed water image (left panel) and detail image (right panel), acquired at 38 keV and high dose.

Figure 3

Examples of linear regressions, performed on high dose images acquired at 32, 35, 38 and 40 keV, from top left panel to bottom right panel. R2 coefficients, slopes and intercepts are given as well.
Examples of linear regressions, performed on high dose images acquired at 32, 35, 38 and 40 keV, from top left panel to bottom right panel. R2 coefficients, slopes and intercepts are given as well.

Noise measurement of the reconstructed attenuation coefficient for water. Relative percentual results are given for μ coefficients, absolute results for the corresponding HU values

Energy (keV)Dosenoiseμ (%)noiseHU
32high4.460
low8.5103
35high3.883
low7.3114
38high3.766
low6.5109
40high3.566
low6.2111

Uniformity of reconstructed attenuation coefficient for water. Relative percentual values are given for μ coefficients, absolute values for the corresponding HU deviations

Energy (keV)Doseuniformityμ (%)uniformityHU
32high1.7917
low3.1932
35high1.4512
low1.4713
38high2.5826
low1.3513
40high2.5424
low1.0710

Low contrast measurements. Relative percentual contrasts are given for μ coefficients, absolute contrasts for the corresponding Hounsfield unit (HU)

Energy (keV)Detail (mm)μ contrast (%)HU contrast
high doselow dosehigh doselow dose
32100.810.7175
80.960.8797
61.100.53104
41.081.121010
20.520.9358
35101.801.851515
81.681.731414
61.561.641313
41.210.87108
20.900.9188
38104.143.864139
83.993.414034
63.664.023641
43.913.453935
23.173.793138
40105.355.495455
85.095.435255
65.124.855249
45.313.415434
25.733.795938

Parameters of the linear regressions between measured CT numbers and tabulated attenuation coefficients

Energy (keV)DoseR2slope (cm–1)intercept (cm–1)
32high0.9996175 ± 40.019 ± 0.009
low0.9996170 ± 40.025 ± 0.009
35high0.9997176 ± 40.018 ± 0.008
low0.9996173 ± 40.017 ± 0.009
38high0.9998177 ± 30.016 ± 0.005
low0.9992175 ± 30.013 ± 0.006
40high0.9992174 ± 70.021 ± 0.012
low0.9991172 ± 70.019 ± 0.012

Accuracies of reconstructed attenuation coefficients for the plastic materials and water, acquired at 38 keV and high dose. Expected values, measured values and relative percentual differences are given for μ coefficients, absolute differences for the corresponding HU deviations

Materialexpected μ (cm–1)measured μ (cm–1)accuracyμ (%)accuracyHU
PE0.2160.221 ± 0.0023.1 ± 1.324 ± 10
BR120.2600.249 ± 0.0044.0 ± 1.637 ± 14
Nylon0.2710.273 ± 0.0030.9 ± 1.08 ± 9
PMMA0.2900.291 ± 0.0030.4 ± 1.24 ± 12
POM0.3590.360 ± 0.0030.2 ± 0.83 ± 11
PTFE0.6220.633 ± 0.0030.1 ± 0.53 ± 11
Water0.2820.278 ± 0.0031.5 ± 0.914 ± 9
eISSN:
1581-3207
Language:
English
Publication timeframe:
4 times per year
Journal Subjects:
Medicine, Clinical Medicine, Radiology, Internal Medicine, Haematology, Oncology