Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems


Cite

Figure 1

Geometry of the measurements, tandem ring applicator and Gafchromic® film located in planes (A), (B) and (C), respectively.
Geometry of the measurements, tandem ring applicator and Gafchromic® film located in planes (A), (B) and (C), respectively.

Figure 2

Digitized dosimetric films as a visual input data for two applicators setup and for all three analyzed dose planes. (I) Applicator ring diameter 26 mm, Gafchromic® film located in planes (A), (B) and (C), respectively. (II) Applicator ring diameter 30 mm: Gafchromic® film located in planes (A), (B) and (C), respectively.
Digitized dosimetric films as a visual input data for two applicators setup and for all three analyzed dose planes. (I) Applicator ring diameter 26 mm, Gafchromic® film located in planes (A), (B) and (C), respectively. (II) Applicator ring diameter 30 mm: Gafchromic® film located in planes (A), (B) and (C), respectively.

Figure 3

Data analysis for the 26 mm ring in plane A. (I)-dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.
Data analysis for the 26 mm ring in plane A. (I)-dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.

Figure 4

Data analysis for the 26 mm ring in plane B. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.
Data analysis for the 26 mm ring in plane B. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.

Figure 5

Data analysis for the 26 mm ring in plane C. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.
Data analysis for the 26 mm ring in plane C. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.

Figure 6

Data analysis for the 30 mm ring in plane A. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.
Data analysis for the 30 mm ring in plane A. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.

Figure 7

Data analysis for the 30 mm ring in plane B. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.
Data analysis for the 30 mm ring in plane B. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.

Figure 8

Data analysis for the 30 mm ring in plane C. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.
Data analysis for the 30 mm ring in plane C. (I) - dose distribution obtained from film, (II) - dose distribution obtained from TPS, (III) - isodoses obtained from film (dots) and TPS (lines), (IV) - gamma factor map.

Results of Gamma analysis for the 26 mm ring and 30 mm ring applicator

Ring, 26 mmGamma averageGamma maxGamma (0.00 – 1.00)
plane A0.221.4299.04%
plane B0.101.2999.31%
plane C0.441.7598.88%
Ring, 30 mmGamma averageGamma maxGamma (0.00 – 1.00)
plane A0.251.9698.11%
plane B0.272.0097.94%
plane C0.101.2299.54%