Effect of the thermoplastic masks on dose distribution in the build-up region for photon beams


Cite

The aim of the study was to investigate the influence of thermoplastic masks material (Klarity Medical&Equipment Co., Guangzhou, China) with different diameters of holes (ϕ 0.25 cm and ϕ 0.40 cm) on the dose distribution in the build-up region for photon beams. Measurements were made for external radiation beams produced by the linear accelerator (TrueBeam, Varian Medical Systems, Inc., Palo Alto, CA, USA) using the Markus parallel plane ionization chamber and the Unidos electrometer (both from PTW, Freiburg, Germany). Measurements were made in a solid water phantom for two photon energies 6 MV and 15 MV, at 90 cm source to skin distance, for four fields of 5 cm × 5 cm, 10 cm × 10 cm, 15 cm × 15 cm and 20 cm × 20 cm. Compared to the open field, the maximum dose with mask was closer to the surface of the phantom by about 1.4 mm and 1.2 mm for 6 MV and 15 MV X-Rays, respectively. The surface dose increase from 10% to 42% for 6 MV and from 5% to 28% for 15 MV X-Rays.