Open Access

Determination of effective source to surface distance and cutout factor in small fields in electron beam radiotherapy: A comparison of different dosimeters


Cite

Objective: The main purpose of this study is to calculate the effective source to surface distance (SSDeff) of small and large electron fields in 10, 15, and 18 MeV energies, and to investigate the effect of SSD on the cutout factor for electron beams a linear accelerator. The accuracy of different dosimeters is also evaluated.

Materials and methods: In the current study, Elekta Precise linear accelerator was used in electron beam energies of 10, 15, and 18 MeV. The measurements were performed in a PTW water phantom (model MP3-M). A Semiflex and Advanced Markus ionization chambers and a Diode E detector were used for dosimetry. SSDeff in 100, 105, 110, 115, and 120 cm SSDs for 1.5 × 1.5 cm2 to 5 × 5 cm2 (small fields) and 6 × 6 cm2 to 20 × 20 cm2 (large fields) field sizes were obtained. The cutout factor was measured for the small fields.

Results: SSDeff in small fields is highly dependent on energy and field size and increases with increasing electron beam energy and field size. For large electron fields, with some exceptions for the 20 × 20 cm2 field, this quantity also increases with energy. The SSDeff was increased with increasing beam energy and field size for all three detectors.

Conclusion: The SSDeff varies significantly for different field sizes or cutouts. It is recommended that SSDeff be determined for each electron beam size or cutout. Selecting an appropriate dosimetry system can have an effect in determining cutout factor.

eISSN:
1898-0309
Language:
English
Publication timeframe:
4 times per year
Journal Subjects:
Medicine, Biomedical Engineering, Physics, Technical and Applied Physics, Medical Physics