The dependence of inhomogeneity correction factors on photon beam quality index performed with the Anisotropic Analytical Algorithm


Cite

Purpose: The purpose of the study was to investigate the dependence of tissue inhomogeneity correction factors (ICFs) on the photon beam quality index (QI).

Materials and Methods: Heterogeneous phantoms, comprising semi-infinite slabs of the lung (0.10, 0.20, 0.26 and 0.30 g/cm3), adipose tissue (0.92 g/cm3) and bone (1.85 g/cm3) in water, were constructed in the Eclipse treatment planning system. Several calculation models of 6 MV and 15 MV photon beams for quality index (TPR20,10) = 0.670±k*0.01 and TPR20,10 = 0.760±k*0.01, k = -3, -2, -1, 0, 1, 2, 3 respectively were built in the Eclipse. The ICFs were calculated with the anisotropic analytical algorithm (AAA) for several beam sizes and points lying at several depths inside of and below inhomogeneities of different thicknesses.

Results: The ICFs increased for lung and adipose tissues with increasing beam quality (TPR20,10), while decreased for bone. Calculations with AAA predict that the maximum difference in ICFs of 1.0% and 2.5% for adipose and bone tissues, respectively. For lung tissue, changes of ICFs of a maximum of 9.2% (6 MV) and 13.8% (15 MV). For points where charged particle equilibrium exists, a linear dependence of ICFs on TPR20,10 was observed. If CPE doesn’t exist, the dependence became more complex. For points inside of the low-density inhomogeneity, the dependence of the ICFs on energy was not linear but the changes of ICFs were smaller than 3.0%. Measurements results carried out with the CIRS phantom were consistent with the calculation results.

Conclusions: A negligible dependence of the ICFs on energy was found for adipose and bone tissue. For lung tissue, in the CPE region, the dependence of ICFs on different beam quality indexes with the same nominal energy may not be neglected, however, this dependence was linear. Where there is no CPE, the dependence of the ICFs on energy was more complicated.

eISSN:
1898-0309
Language:
English
Publication timeframe:
4 times per year
Journal Subjects:
Medicine, Biomedical Engineering, Physics, Technical and Applied Physics, Medical Physics