Open Access

A numerical study on the effect of conductivity change in cell kill distribution in irreversible electroporation

   | Jun 25, 2020

Cite

Introduction: irreversible electroporation (IRE) is a tissue ablation technique and physical process used to kill the undesirable cells. In the IRE process by mathematical modelling we can calculate the cell kill probability and distribution inside the tissue. The purpose of the study is to determine the influence of electric conductivity change in the IRE process into the cell kill probability and distribution.

Methods: cell death probability and electric conductivity were calculated with COMSOL Multiphysics software package. 8 pulses with a frequency of 1 Hz, pulse width of 100 µs and electric field intensity from 1000 to 3000 V/Cm with steps of 500 V/Cm used as electric pulses.

Results: significantly, the electrical conductivity of tissue will increase during the time of pulse delivery. According to our results, electrical conductivity increased with an electric field intensity of pulses. By considering the effect of conductivity change on cell kill probability, the cell kill probability and distribution will change.

Conclusion: we believe that considering the impact of electric conductivity change on the cell kill probability will improve the accuracy of treatment outcome in the clinic for treatment with IRE.

eISSN:
1898-0309
Language:
English
Publication timeframe:
4 times per year
Journal Subjects:
Medicine, Biomedical Engineering, Physics, Technical and Applied Physics, Medical Physics