Cite

Aim: To conduct a study on the effect of random setup errors inpatient for dose delivery in Intensity Modulated Radiotherapy plans using Octavius 4D phantom.

Materials and methods: 11 patients with cancer of H&N were selected for this study. An IMRT plan was created for each patient. The IMRT quality assurance plans were transferred to Mosaiq workstation in a linear accelerator. These plans were delivered at the reference treatment position. Subsequently, the QA plans were delivered on the Octavius 4D phantom after introducing errors in various translational and rotational directions. The setup inaccuracies introduced varied from 1 mm to 5 mm along X, Y. These setup uncertainties were then introduced along X and Y direction simultaneously in equal measures. Similarly, IMRT plans were delivered also after introducing roll and yaw rotation of 1, 2 and 3 degrees in phantom. The deviation of gamma indices at all these positions was analyzed with respect to the reference setup position.

Results: The percentage of points passing the gamma acceptance criterion decrease as we increase the setup error. The change is found to be very insignificant with setup error up to 2 mm along X, Y or XY direction. Similarly, the rotational error of up to 3 degrees is found to be acceptable.

Conclusions: Small setup (< 2 mm) correction in patients may not adversely affect the dose delivery. But an error of similar magnitude in 2 directions simultaneously has a much greater impact on IMRT dose delivery.

eISSN:
1898-0309
Language:
English
Publication timeframe:
4 times per year
Journal Subjects:
Medicine, Biomedical Engineering, Physics, Technical and Applied Physics, Medical Physics