Interpretation of Gamma Index for Quality Assurance of Simultaneously Integrated Boost (SIB) IMRT Plans for Head and Neck Carcinoma


Cite

Objective: The Gamma Index is prerequisite to estimate point-by-point difference between measured and calculated dose distribution in terms of both Distance to Agreement (DTA) and Dose Difference (DD). This study aims to inquire what percentage of pixels passing a certain criteria assure a good quality plan and suggest gamma index as efficient mechanism for dose verification of Simultaneous Integrated Boost Intensity Modulated Radiotherapy plans.

Method: In this study, dose was calculated for 14 head and neck patients and IMRT Quality Assurance was performed with portal dosimetry using the Eclipse treatment planning system. Eclipse software has a Gamma analysis function to compare measured and calculated dose distribution. Plans of this study were deemed acceptable when passing rate was 95% using tolerance for Distance to agreement (DTA) as 3mm and Dose Difference (DD) as 5%.

Result and Conclusion: Thirteen cases pass tolerance criteria of 95% set by our institution. Confidence Limit for DD is 9.3% and for gamma criteria our local CL came out to be 2.0% (i.e., 98.0% passing). Lack of correlation was found between DD and γ passing rate with R2 of 0.0509. Our findings underline the importance of gamma analysis method to predict the quality of dose calculation. Passing rate of 95% is achieved in 93% of cases which is adequate level of accuracy for analyzed plans thus assuring the robustness of SIB IMRT treatment technique. This study can be extended to investigate gamma criteria of 5%/3mm for different tumor localities and to explore confidence limit on target volumes of small extent and simple geometry.

eISSN:
1898-0309
Language:
English
Publication timeframe:
4 times per year
Journal Subjects:
Medicine, Biomedical Engineering, Physics, Technical and Applied Physics, Medical Physics