Photon beam commissioning of an Elekta Synergy linear accelerator

Open access

Abstract

The aim of this study is to present the results of commissioning of Elekta Synergy linear accelerator (linac). The acceptance test and commissioning were performed for three photon beams energies 4 MV, 6 MV and 15 MV and for the multileaf collimator (MLC). The percent depth doses (PDDs), in-plane and cross-plane beam profiles, head scatter factors (Sc), relative photon output factors (Scp), universal wedge transmission factor and MLC transmission factors were measured. The size of gantry, collimator, and couch isocenter were also measured.

If the inline PDF is not rendering correctly, you can download the PDF file here.

  • [1] Narayanasamy G Saenz D Cruz W et al. Commissioning an Elekta Versa HD linear accelerator. J Appl Clin Med Phys. 2016;17(1):179-191.

  • [2] Allen J. White Paper: High Dose Rate mode (Flattening Filter Free) Radiotherapy Clinical advantages of High Dose Rate mode available with Agility™ on Elekta’s Versa HD™ linear accelerator [white paper]. 2014.

  • [3] Boda-Heggemann J Mai S Fleckenstein J et al. Flattening-filter-free intensity modulated breath-hold image-guided SABR (Stereotactic ABlative Radiotherapy) can be applied in a 15-min treatment slot. Radiother Oncol. 2013;109(3):505-509.

  • [4] Das IJ Cheng CW Watts RJ et al. Accelerator beam data commissioning equipment and procedures: report of the TG-106 of the Therapy Physics Committee of the AAPM. Med Phys. 2008;35(9):4186-4215.

  • [5] Klein EE Hanley J Bayouth J et al. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36(9):4197-4212.

  • [6] Almond PR Biggs PJ Coursey BM et al. AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys. 1999;26(9):1847-1870.

  • [7] Niroomand-Rad A Blackwell CR Coursey BM et al. Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. Med Phys. 1998;25(11):2093-2115.

  • [8] International Electrotechnical Commission. Medical electrical equipment. Part 2: Particular requirements for the safety of gamma beam therapy equipment. IEC 60601-2-11. Geneva Switzerland: IEC; 1987.

  • [9] Pönisch F Titt U Vassiliev ON et al. Properties of unflattened photon beams shaped by a multileaf collimator. Med Phys. 2006;33(6):1738-1746.

  • [10] Stathakis S Esquivel C Gutierrez A et al. Treatment planning and delivery of IMRT using 6 and 18 MV photon beams without flattening filter. Appl Radiat Isot. 2009;67(9):1629-1637.

  • [11] Weber L Nilsson P Ahnesjö A. Build-up cap materials for measurement of photon head-scatter factors. Phys Med Biol. 1997;42(10):1875-1886.

  • [12] Jursinic PA. Measurement of head scatter factors of linear accelerators with columnar miniphantoms. Med Phys. 2006;33(6):1720-1728.

  • [13] Dieterich S Sherouse GW. Experimental comparison of seven commercial dosimetry diodes for measurement of stereotactic radiosurgery cone factors. Med Phys. 2011;38(7):4166-4173.

  • [14] Phillips MH Parsaei H Cho PS. Dynamic and omni wedge implementation on an Elekta SL linac. Med Phys. 2000;27(7):1623-1634.

  • [15] Thompson CM Weston SJ Cosgrove VC Thwaites DI. A dosimetric characterization of a novel linear accelerator collimator. Med Phys. 2014;41(3):031713.

  • [16] International Electrotechnical Commission. Medical electrical equipment — Part 2.1. Particular requirements for the basic safety and essential performance of electron accelerators in the range 1 MeV to 50 MeV. IEC 60601–2-1 Ed. 3.0. Geneva Switzerland: IEC; 2009.

  • [17] Chang Z Wu Q Adamson J et al. Commissioning and dosimetric characteristics of TrueBeam system: composite data of three TrueBeam machines. Med Phys. 2012;39(11):6981-7018.

  • [18] Kragl G af Wetterstedt S Knäusl B et al. Dosimetric characteristics of 6 and 10 MV unflattened photon beams. Radiother Oncol. 2009;93(1):141-146.

  • [19] Benedict SH Yenice KM Followill D et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys. 2010;37(8):4078-4101.

  • [20] Glide-Hurst C Bellon M Foster R et al. Commissioning of the Varian TrueBeam linear accelerator: a multi-institutional study. Med Phys. 2013;40(3):031719.

  • [21] Cashmore J. The characterization of unflattened photon beams from a 6 MV linear accelerator. Phys Med Biol. 2008;53(7):1933-1346.

  • [22] Vassiliev ON Titt U Pönisch F et al. Dosimetric properties of photon beams from a flattening filter free clinical accelerator. Phys Med Biol. 2006;51(7):1907-1917.

Search
Journal information
Impact Factor


CiteScore 2018: 0.38

ICV 2017 = 103.49

SCImago Journal Rank (SJR) 2018: 0.132
Source Normalized Impact per Paper (SNIP) 2018: 0.303

Metrics
All Time Past Year Past 30 Days
Abstract Views 0 0 0
Full Text Views 782 362 11
PDF Downloads 412 220 6