Determination of contributions of scatter and distance error to the source strength of 192Ir HDR brachytherapy source

Open access

Abstract

High dose rate (HDR) brachytherapy commonly employs a 192Ir encapsulated source to deliver high dose to the malignant tissues. Calibrations of brachytherapy sources are performed by the manufacturer using a well-type chamber or by in-air measurement using a cylindrical ionization chamber. Calibration using the latter involves measurements to be carried out at several distances and room scatter can also be determined. The aim of the present study is to estimate the scatter contribution from the walls, floor and various materials in the room in order to determine the reference air kerma rate of an 192Ir HDR brachytherapy source by in-air measurements and also to evaluate the error in the setup distance between the source centre and chamber centre. Air kerma measurements were performed at multiple distances from 10 cm to 40 cm between the source and chamber. The room scatter correction factor was determined using the iterative technique. The distance error of −0.094 cm and −0.112 cm was observed for chamber with and without buildup cap respectively. The scatter component ranges from 0.3% to 5.4% for the chamber with buildup cap and 0.3% to 4.6% without buildup cap for distances between 10 to 40 cm respectively. Since the average of the results at multiple distances is considered to obtain the actual air kerma rate of the HDR source, the seven distance method and iterative technique are very effective in determining the scatter contribution and the error in the distance measurements.

References

  • [1] Chang L, Ho SY, Chui CS, et al. Room scatter factor modelling and measurement error analysis of 192Ir HDR calibration by a Farmer chamber. Phys Med Biol. 2007;52(3):871-877.

  • [2] AAPM. Specification of brachytherapy source strength. New York, NY: American Association of Physicists in Medicine, AAPM Report 21; 1987.

  • [3] Nath R, Anderson LL, Luxton G, et al. Dosimetry of interstitial brachytherapy sources. Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. Med Phys. 1995;22(2):209-234.

  • [4] Goetsch SJ, Attix FH, Pearson DW, et al. Calibration of 192Ir high dose rate afterloading systems. Med Phys. 1991;18(3):462-467.

  • [5] Stump KE, DeWerd LA, Micka JA, et al. Calibration of new high dose rate 192Ir sources. Med Phys. 2002;29(7):1483-1488.

  • [6] Rasmussen BE, Davis SD, Schmidt CR, et al. Comparison of air-kerma strength determinations for HDR 192Ir sources. Med Phys. 2011;38(12):6721-6729.

  • [7] Palani Selvam T, Govinda Rajan KN, Nagarajan PS, et al. Monte Carlo aided room scatter studies in the primary air kerma strength standardization of a remote afterloading 192Ir HDR source. Phys Med Biol. 2001;46(9):2299-2315.

  • [8] Patel NP, Majumdar B, Vijayan V. Study of scattered radiation for in-air calibration by a multiple-distance method using ionization chambers and an HDR 192Ir brachytherapy source. Br J Radiol. 2006;79(940):347-352.

  • [9] Patel NP, Majumdar B, Vijiyan V, et al. In-air calibration of an HDR 192Ir brachytherapy source using therapy ion chambers. J Cancer Res Ther. 2005;1(4):213-220.

  • [10] Kumar S, Srinivasan P, Sharma SD, et al. Evaluation of scatter contribution and distance error by iterative methods for strength determination of HDR 192Ir brachytherapy source. Med Dos. 2010;35(3):230-237.

  • [11] IAEA. Calibration of photon and beta ray sources used in brachytherapy - Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and hospitals. Vienna: International Atomic Energy Agency, IAEA-TECDOC-1274; 2002.

  • [12] Kondo S, Randolph ML. Effect of finite size of ionization chambers on measurements of small photon sources. Rad Res. 1960;13(1):37-60.

  • [13] Bielajew AF. Correction factors for thick-walled ionization chambers in point source photon beams. Phys Med Biol. 1990;35(4):501-516.

  • [14] Deutsche Gesellschaft für Medizinische Physik e.V. Praktische Dosimetrie in der HDR-Brachytherapie. DGMP-Bericht 13. 1999.

  • [15] Deutsches Institut fuer Normung. DIN 6809, Teil 2, Klinische Dosimetrie, Brachytherapie mit umschlossenen gammastrahlenden radioaktiven Stoffen. Berlin: Beuth Verlag; 1993.

  • [16] Baltas D, Geramani K, Ioannidis GT, et al. Comparison of calibration procedures for 192Ir high-dose-rate brachytherapy sources. Int J Radiat Oncol Biol Phys. 1999;43(3):653-661.

  • [17] Marechal MH, de Almeida CE, Ferreira IH, et al. Experimental derivation of wall correction factors for ionization chambers used in high dose rate 192Ir source calibration. Med Phys. 2002;29(1):1-5.

  • [18] IAEA. Absorbed Dose Determination in Photon and Electron Beams, An International Code of Practice. Vienna: International Atomic Energy Agency, Technical Reports Series No. 277 (2nd Edition); 1987.

Polish Journal of Medical Physics and Engineering

The Journal of Polish Society of Medical Physics

Journal Information


CiteScore 2016: 0.08
ICV 2016 = 98.75

SCImago Journal Rank (SJR) 2016: 0.112
Source Normalized Impact per Paper (SNIP) 2016: 0.031

Metrics

All Time Past Year Past 30 Days
Abstract Views 0 0 0
Full Text Views 17 17 15
PDF Downloads 2 2 2