A Fast-Multi-Channel Sub-Millikelvin Precision Resistance Thermometer Readout Based on the Round-Robin Structure

Open access

Abstract

The fast response multipoint high-precision temperature measurement is often necessary in many dynamical measurement fields and industrial applications. However, limited by the existing electric circuit architecture, either the AC or DC bridges have the shortcoming that the rates or precisions degenerate markedly in the multi-channel scanning mode. To overcome this disadvantage, a round-robin structural low-cost ratiometric resistance thermometer readout based on several commercial 32-bit sigma-delta analogue-to-digital converters (Σ-Δ ADCs) was presented in this article. The experimental results show that the precision of this readout corresponds to 0.1 mK at 1 Hz when sampling four channel resistors simultaneously, while the precision and rate are not degenerating with the channel number increasing. In addition, the uncertainty of the readout is investigated in this article. It shows that the presented readout can achieve an uncertainty as low as 2.1 mK at 1 Hz (K = 2).

[1] Frohlich, T., Fehling, T., Heydenbluth, D. (2009). Mass dissemination using a robot system. tm - Technisches Messen, 76 (7-8), 382-387.

[2] Schweiger, H.G., Multerer, M., Gores, H.J. (2007). Fast multichannel precision thermometer. IEEE Transactions on Instrumentation and Measurement, 56 (5), 2002-2009.

[3] Eke, R., Kavasoglu, A.S., Kavasoglu, N. (2012). Design and implementation of a low-cost multi-channel temperature measurement system for photovoltaic modules. Measurement, 45, 1499-1509.

[4] Pearce, J.V., Gray, J., Veltcheva, R.I. (2016). Characterisation of a selection of AC and DC resistance bridges for platinum resistance thermometry. International Journal of Thermophysics, 37, 109.

[5] Smorgon, D., Fernicola, V.C., Coslovi, I. (2011). Lowcost ratiometric front-end for industrial PRT applications. International Journal of Thermophysics, 32, 2317-2324.

[6] Ambrosetti, R., Matteoli, E., Ricci, D. (2012). Note: A versatile, stable, high-resolution readout system for RTD and thermistor sensors. Review of Scientific Instruments, 83, 096101.

[7] Taylor, H.R., Navarro, H.A. (1983). A microcomputerbased instrument for applications in platinum resistance thermometry. Journal of Physics E: Scientific Instruments, 16, 1100-1104.

[8] Bramley, P., Pickering, J.R. (2006). Better accuracy in temperature calibration and measurement through a new type analog-to-digital converter. Cal Lab Magazine, 10/11, 21-26.

[9] Walker, R., Willgress, N. (2003). Achieving 0.25 mK uncertainty with an integrated-circuit resistance thermometer readout. In XVII IMEKO World Congress, Dubrovnik, Croatia, 1594-1597.

[10] White, D.R., Williams, J.M., Ramsey I.E. (1997). A simple resistance network for the calibration of resistance bridges. IEEE Transactions on Instrumentation and Measurement, 42 (5), 1068-1074.

[11] White, D.R., Clarkson, M.T., Saunders, P., Yoon, H. (2008). A general technique for calibrating indicating instruments. Metrologia, 45, 199-210.

[12] Analog Devices, Inc. (2015-2016). 32-bit, 10 kSPS, Sigma-Delta ADC with 100 μs settling and true rail-torail buffers. Data Sheet AD7177-2. http://www.analog.com/media/en/technical-documentation/datasheets/AD7177-2.pdf.

[13] Vishay Precision Group. (2014). Hermetically sealed miniature ultra high precision Z-foil technology resistors with TCR of 0.05 ppm/°C, tolerance of ± 0.001% and load life stability of ± 0.005 %, unaffected by humidity. VHP203 (Z-Foil). http://www.vishaypg.com/docs/63146/vhp203.pdf.

[14] Analog Devices, Inc. (2013-2015). High performance, low power, rail-to-rail precision instrumentation amplifier. Data Sheet AD8422. http://www.analog.com/media/en/technical-documentation/datasheets/AD8422.pdf.

[15] Analog Devices, Inc. (2005-2017). 0.4Ω CMOS, dual DPDT switch in WLCSP/LFCSP/TSSOP. Data Sheet ADG888. http://www.analog.com/media/en/technicaldocumentation/data-sheets/ADG888.pdf.

[16] Nicholas, J.V., White, D.R. (2001). Traceable Temperatures: An Introduction to Temperature Measurement and Calibration. John Wiley & Sons.

[17] Walker, R. (2010). Recent advances in resistance thermometry readouts. In NCSL International Workshop and Symposium.

[18] Palencar, R., Sopkuliak, P., Palencar, J., Duris, S., Suroviak, E., Halaj, M. (2017). Application of Monte Carlo method for evaluation of uncertainties of ITS-90 by standard platinum resistance thermometer. Measurement Science Review, 17 (3), 108-116.

[19] Duris, S., Palencar, R., Ranostaj, J. (2008). Contribution of the SPRT calibration to uncertainty of temperature T-90 measured by the calibrated SPRT. Measurement Science Review, 8 (1), 5-10.

[20] Walker, R. (2011). Automatic linearity calibration in a resistance thermometry bridge. International Journal of Thermophysics, 32 (1-2), 215-223.

Measurement Science Review

The Journal of Institute of Measurement Science of Slovak Academy of Sciences

Journal Information


IMPACT FACTOR 2017: 1.345
5-year IMPACT FACTOR: 1.253



CiteScore 2017: 1.61

SCImago Journal Rank (SJR) 2017: 0.441
Source Normalized Impact per Paper (SNIP) 2017: 0.936

Metrics

All Time Past Year Past 30 Days
Abstract Views 0 0 0
Full Text Views 34 34 10
PDF Downloads 29 29 9