Numerical Analysis of a Flexible Dual Loop Coil and its Experimental Validation for pre-Clinical Magnetic Resonance Imaging of Rodents at 7 T

Abstract

A surface radio frequency coil was developed for small animal image acquisition in a pre-clinical magnetic resonance imaging system at 7 T. A flexible coil composed of two circular loops was developed to closely cover the object to be imaged. Electromagnetic numerical simulations were performed to evaluate its performance before the coil construction. An analytical expression of the mutual inductance for the two circular loops as a function of the separation between them was derived and used to validate the simulations. The RF coil is composed of two circular loops with a 5 cm external diameter and was tuned to 300 MHz and 50 Ohms matched. The angle between the loops was varied and the Q factor was obtained from the S11 simulations for each angle. B1 homogeneity was also evaluated using the electromagnetic simulations. The coil prototype was designed and built considering the numerical simulation results. To show the feasibility of the coil and its performance, saline-solution phantom images were acquired. A correlation of the simulations and imaging experimental results was conducted showing a concordance of 0.88 for the B1 field. The best coil performance was obtained at the 90° aperture angle. A more realistic phantom was also built using a formaldehyde-fixed rat phantom for ex vivo imaging experiments. All images showed a good image quality revealing clearly defined anatomical details of an ex vivo rat.

If the inline PDF is not rendering correctly, you can download the PDF file here.

  • [1] Kiesslin, F., Pichler, B.J. (2011). Small Animal Imaging: Basics and Practical Guide. Springer.

  • [2] Darrasse, L., Ginefri, J.C. (2003). Perspectives with cryogenic RF probes in biomedical MRI. Biochimie, 85 (9), 915-937.

  • [3] Childs, A.S., Malik, S.J., O’Regan, D.P., Hajnal, J.V. (2013). Impact of number of channels on RF shimming at 3T. Magnetic Resonance Materials in Physics, Biology and Medicine, 26 (4), 401-410.

  • [4] Doty, F.D., Entzminger, G., Kulkarni, J., Pamarthy, K., Staab, J.P. (2007). Radio frequency coil technology for small-animal MRI. NMR in Biomedicine, 20 (3), 304-325.

  • [5] Fujita, H. (2007). New horizons in MR technology: RF coil designs and trends. Magnetic Resonance in Medical Sciences, 6 (1) 29-42.

  • [6] Gilbert, K.M., Curtis, A.T., Gati, J.S., Martyn Klassen, L., Villemaire, L.E., Menon, R.S. (2010). Transmit/receive radiofrequency coil with individually shielded elements. Magnetic Resonance in Medicine, 64 (6), 1640-1651.

  • [7] Hayes, C.E. (2009). The development of the birdcage resonator: A historical perspective. NMR in Biomedicine, 22 (9), 908-918.

  • [8] Novikov, A. (2011). Advanced theory of driven birdcage resonator with losses for biomedical magnetic resonance imaging and spectroscopy. Magnetic Resonance Imaging, 29 (2), 260-271.

  • [9] Shajan, G., Hoffmann, J., Balla, D., Pohmann, R. (2010). A 700MHz receive array using patch antenna for spin excitation. In Proceedings of the 18th Annual Meeting of ISMRM, May 1-7, 2010, Stockholm, Sweden.

  • [10] Vaughan, J.T., Griffiths, J.R. (2012). RF Coils for MRI. John Wiley & Sons.

  • [11] Xin, X., Han, J., Feng, Y., Feng, Q., Chen, W. (2012). Inverse design of an organ-oriented RF coil for open, vertical-field, MR-guided, focused ultrasound surgery. Magnetic Resonance Imaging, 30 (10), 1519-1526.

  • [12] Jin, J. (1998). Electromagnetic Analysis and Design in Magnetic Resonance Imaging. CRC Press.

  • [13] Kumar, A., Bottomley, P.A. (2008). Optimized quadrature surface coil designs. Magnetic Resonance Materials in Physics, Biology and Medicine, 21 (1-2), 41-52.

  • [14] Constantinides, C., Angeli, S. (2012). Elimination of mutual inductance in NMR phased arrays: The paddle design revisited. Journal of Magnetic Resonance, (222), 59-67.

  • [15] Stara, R., Fontana, N., Alecci, M., Alfonsetti, M., Galante, A.,Vitacolonna, A., Biagi, L., Buonincontri, G., Del Guerra, A., Tosetti, M., Manara, G., Monorchio, A., Tiberi, G., Retico, A. (2011). RF coil design for low and high field MRI: Numerical methods and measurements. In 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), October 23-29, 2011. IEEE, 3465-3469.

  • [16] Weiland, T. (1996). Time domain electromagnetic field computation with finite difference methods. International Journal of Numerical Modelling: Electronic Networks, Devices and Fields, 9 (4), 295-319.

  • [17] Solis-Najera, S.E., Martin, R., Vazquez, F., Rodriguez, A.O. (2015). Surface coil with reduced specific absorption rate for rat MRI at 7 T. Magnetic Resonance Materials in Physics, Biology and Medicine, 28, 599-608.

  • [18] Berenger, J.-P. (1996). Three-dimensional perfectly matched layer for the absorption of electromagnetic waves. Journal of Computational Physics, 127 (2), 363-379.

  • [19] Peterson, A.F., Ray, S.L., Mittra, R. (1997). Computational Methods for Electromagnetics, 1st Ed. Wiley-IEEE Press.

  • [20] Gabriel, S., Lau, R.W., Gabriel, C. (1996). The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. Physics in Medicine and Biology, 41 (11), 2251-2269.

OPEN ACCESS

Journal + Issues

Search