A High Pressure Resonator Transducer with a Programmed Correction of Static Characteristics

Open access

Abstract

A high pressure resonator transducer (0 to 100 MPa) devised by the author has been described. The elastic element of the converter consists of a cylinder with an offset arranged axis hole. Quartz resonators were used for the measurement of deformations of the pipe. Based upon the results of the transducer testing, a new algorithmic method for the minimizalizsation of the temperature error, that eliminates the need for a temperature gauge has been worked out.

[1] Newhall, D. H., Abbot, L. H. (1961). High-Pressure Measurement. Instruments & Control Systems.

[2] Wisniewski, R., Molinar, G. (1999). Pressure standards and sensors up to 3 GPa, actual state and developmenttrends. Physica B: Condensed Matter, 265(1-4), 246−253.

[3] Gillum, D. R. (2009). Industrial Pressure, Level, and Density Measurement (2nd Edition). International Societyof Automation.

[4] Kobata, T. (2005). Characterization of quartz Bourdon-type high-pressure transducers. Metrologia, 42(6), S235S238.

[5] Kocas, I., Bergoglio, M. (December 2012). An investigation of quartz type pressure transducer behavior undercontinuous pressure conditions and metrological characterization. Measurement: Journal of theInternational Measurement Confederation, 45(10), 2486−2489.

[6] Jia, R. (1999). Principle of pressure-temperature transducer of quartz crystal resonator. Proceedings of theAnnual IEEE International Frequency Control Symposium, 1054−1057.

[7] EerNisse, E. P. (1987). Theoretical modeling of quartz resonator pressure transducers, Proceedings of the AnnualIEEE International Frequency Control Symposium, 339−343.

[8] Werszko, M., Werszko, R. Rezonatorowy przetwornik cisnienia (A resonator pressure transducer), Patent PL 178676.

[9] Karrer, E., Ward, R. (1977). A low-range quartz resonator pressure transducer, ISA Transactions, 16(2), 90−98.

[10] Cheshmehdoost, A., Jones, B. E. (June 1995). Design and performance characteristics of an integrated highcapacity DETF-based force sensor, The 8th International Conference on Solid State Sensors andActuators, and Eurosensors IX, Stockholm, Sweden, 52(1-3), 99−102.

[11] Bryzek, J. (July 1994). Signal conditioning for smart sensors and transducers, Sensors (Peterborough, NH),11(7), 151−160.

[12] Gleser, L. J. (1998). Assessing uncertainty in measurement, Statistical Science, 13(3).

[13] Yadav, S., Gupta, V.K., Bandyopadhyay, A.K. (2010). Investigations on measurement uncertainty and stabilityof pressure dial gauges and transducers, Measurement Science Review, 10(4), 130−135.

Metrology and Measurement Systems

The Journal of Committee on Metrology and Scientific Instrumentation of Polish Academy of Sciences

Journal Information


IMPACT FACTOR 2016: 1.598

CiteScore 2016: 1.58

SCImago Journal Rank (SJR) 2016: 0.460
Source Normalized Impact per Paper (SNIP) 2016: 1.228

Cited By

Metrics

All Time Past Year Past 30 Days
Abstract Views 0 0 0
Full Text Views 72 72 6
PDF Downloads 27 27 2