Open Access

Analytical procedure for the determination of tetracyclines in medicated feedingstuffs by liquid chromatography-mass spectrometry


Cite

Introduction: The article presents a rapid and simple analytical procedure for determination of four tetracyclines (oxytetracycline, tetracycline, chlortetracycline, and doxycycline) in animal medicated feedingstuffs. Material and Methods: Two-gramme samples were extracted by a Na2EDTA-McIlvaine buffer (pH 4)/methanol mixtures (40/60, v/v). The determination was achieved by liquid chromatography using a Zorbax Eclipse XDB C18 analytical column with mass spectrometer detection (LC-MS). Results: Recoveries of the antibiotics from spiked feed samples ranged from 78.2% to 113.5%. The LOD and LOQ for tetracyclines in feeds ranged from 2.8 to 4.2 and from 4.3 to 5.7 mg/kg, respectively. Conclusion: The method was successfully validated and proved to be efficient, precise, and useful for quantification of tetracyclines in medicated feedingstuffs.

eISSN:
2450-8608
Language:
English
Publication timeframe:
4 times per year
Journal Subjects:
Life Sciences, Molecular Biology, Microbiology and Virology, other, Medicine, Veterinary Medicine