Open Access

Performance tests using the Lexsyg luminescence reader


Cite

This study presents test measurements on a Lexsyg luminescence reader, acquired by the Giessen luminescence group. The reader is of the type Standard, hence designed for routine determination of palaeodoses using quartz or K-feldspars. The tests include measurements of the stimulation powers, preheat temperatures, OSL- and TL-curves, as well as dose recovery tests and calibration measurements using highly sensitised calibration quartz. A comparison of D e values determined using a Lexsyg reader with single grain D e-values, which have been previously obtained on a Risø reader, is also presented. The results imply that the Lexsyg reader is a highly reliable measurement device with high reproducibility, yielding instrumental uncertainties of around 0.3%.

eISSN:
1897-1695
Language:
English
Publication timeframe:
Volume Open
Journal Subjects:
Geosciences, other