Survey of Equipment Quality Control in Radiotherapy Centres in Croatia: First Results

Open access

Survey of Equipment Quality Control in Radiotherapy Centres in Croatia: First Results

Implementation of advanced radiation therapy techniques in clinical practice can greatly improve tumour control and normal tissue sparing. An important part of this implementation is quality control (QC) of every part of the radiotherapy process, as it helps to detect errors and provides instant remedy. This increases the probability of successful radiation treatment and ensures patient radiation safety. Every radiotherapy quality assurance (QA) programme is based on quality control of radiotherapy equipment. The aim of our survey was to review QC practices in a number of radiotherapy centres in Croatia. As a first step, we defined a set of tests to check different parameters of linear accelerators and simulators in these centres. The tests were defined and performed according to protocols developed at two university hospitals. Test results varied largely between the centres. This calls for harmonisation of QC protocols.

If the inline PDF is not rendering correctly, you can download the PDF file here.

  • Meijer GJ van Kleffens HJ Mijnheer BJ. Consistency in quality control programmes for electron accelerators in radiotherapy centres. Radiother Oncol 1998;48:103-10.

  • Švabić M Jurković S Faj D Kasabasić M Smilović Radojčić D Ivković A. Practises of radiotherapy equipment quality control in radiotherapy centres in Croatia. Coll Antropol 2008;32(Suppl 2):217-9.

  • Klein EE Hanley J Bayouth J Yin FF Simon W Dresser S Serago C Aguirre F Ma L Arjomandy B Liu C Sandin C Holmes T; Task Group 142 American Association of Physicists in Medicine. Quality assurance of medical accelerators. Med Phys 2009;36:4197-212.

  • Nath R Biggs PJ Bova FJ Ling CC Purdy JA van de Geijn J Weinhous MS. AAPM code of practice for radiotherapy accelerators: of AAPM Radiation Therapy Task Group No. 45. Med Phys 1994;21:1093-121.

  • International Atomic Energy Agency (IAEA). Comprehensive audits of radiotherapy practices: a tool for quality improvement. Vienna: IAEA; 2007.

  • Croatian Medical and Biological Engineering Society / Medical Physics Division. Priručnik za kontrolu kvalitete linearnog akceleratora [Linear accelerator quality control handbook in Croatian] [displayed 25 July 2011]. Available at http://medfizika.netai.net/upload/Protokol_OncorOsijekRijeka.doc

  • Van Dam J Johansson KA Bridier A Sernbo G Hansson U. EORTC radiotherapy group quality assurance: mechanical checks and beam alignments of megavoltage equipment. Radiother Oncol 1993;29:91-6.

  • Rajapakshe R Luchka KB Shalev S. A quality control test for electronic portal imaging devices. Med Phys 1996;23:1237-44.

  • Droege RT. A practical method to routinely monitor resolution in digital images. Med Phys 1983;10:337-43.

  • Podgorsak EB. Radiation Oncology Physics: A Handbook for Teachers and Students. Vienna: IAEA; 2005.

Search
Journal information
Impact Factor

IMPACT FACTOR 2018: 1.436
5-year IMPACT FACTOR: 1.606

CiteScore 2018: 1.53

SCImago Journal Rank (SJR) 2018: 0.358
Source Normalized Impact per Paper (SNIP) 2018: 0.608

Metrics
All Time Past Year Past 30 Days
Abstract Views 0 0 0
Full Text Views 228 107 5
PDF Downloads 95 64 4