Determination of phosphate in water by flow coulometry


Cite

In this work, the possibility of flow coulometry application as coulometric titration was studied. The method was used to analyze phosphates in wastewater samples. The principle of the determination consisted in the formation of molybdophosphate and its subsequent one-electron electrolytic reduction. The present method is applicable under optimal conditions in the concentration range of 1.5 × 10−6 to 5.5 × 10−5 mol/dm3. Detection limit of the method is 3.42 × 10−7 mol/dm3. Mineralization step has been proved a problem in total phosphate content determination. If mineralization was not carried out, only inorganic soluble phosphates were determined. It is a new method characterized by its simplicity of instrumentation and handling, which is a prerequisite for its further use in the field of trace analysis.