Search Results

You are looking at 1 - 1 of 1 items for

  • Author: Éva Katalin Kelemen x
Clear All Modify Search
Open access

Béla Kovács, Lajos Kristóf Kántor, Mircea Dumitru Croitoru, Éva Katalin Kelemen, Mona Obreja, Előd Ernő Nagy, Blanka Székely-Szentmiklósi and Árpád Gyéresi

Abstract

A reverse-phase HPLC (RP-HPLC) method was developed for strontium ranelate using a full factorial, screening experimental design. The analytical procedure was validated according to international guidelines for linearity, selectivity, sensitivity, accuracy and precision. A separate experimental design was used to demonstrate the robustness of the method. Strontium ranelate was eluted at 4.4 minutes and showed no interference with the excipients used in the formulation, at 321 nm. The method is linear in the range of 20–320 μg mL−1 (R2 = 0.99998). Recovery, tested in the range of 40–120 μg mL−1, was found to be 96.1–102.1 %. Intra-day and intermediate precision RSDs ranged from 1.0–1.4 and 1.2–1.4 %, resp. The limit of detection and limit of quantitation were 0.06 and 0.20 μg mL−1, resp. The proposed technique is fast, cost-effective, reliable and reproducible, and is proposed for the routine analysis of strontium ranelate.